40
Document No. 896 623
3.10.2
Scavenging Through Reservoir
SAMPLE GAS OUT
Figure 11
Scavenging through Ventilator Reservoir
•
Connect an exhaust line to the sample gas outlet on the module’s front panel.
•
Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube
diameter is at least 2-3 times larger than the exhaust line.
3.10.3
Scavenging Through Direct Connection
SAMPLE GAS OUT
Figure 12
Connection Directly to a Scavenging System
•
Connect a T-fitting between the monitor and the exhaust line.
•
Attach the other end of the line to the scavenging tube.