General Service Guide
17
Document No. 896 623
3.3.3
Connection to mains
Connect the power cord to the mains power inlet on the side of the monitor and to the wall socket.
WARNING
The power cord may only be connected to a three-wire, grounded, hospital grade
receptacle.
3.3.4
Connection to Network
To connect your monitor to Datex-Ohmeda Network, make sure you have a network board B-NET or
B-UPINET installed.
Use the Monitor-Network cable to connect the monitor to the network.
1.
Make sure that the power is turned off.
2.
Connect the one of RJ-45 connectors to the connector and the Identification Plug to X2
connector on the network board.
3.
Connect the other RJ-45 connector to the corresponding connector on the wallbox.
4.
Turn on the power. Confirm that the network symbol and 'Connected to Network' message
are displayed on the upper part of the screen.
NOTE:
The network symbol will not be displayed if the battery symbol is displayed on the upper part
of the screen.
3.3.5
Inserting the parameter modules
Insert each plug-in parameter module into a module slot. Firmly press the module in position.
NOTE:
Do not use two or more modules with identical functions in the monitor. Take special care of
this when using the extension frame.
Modules with identical functions are:
•
Haemodynamic multiparameter modules: M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-
NETPR, M-NESTR, M-NE12STPR, M-NE12TPR and M-NE12STR
•
Pressure Modules: M-P, Pressure Temp Module, M-PT
•
Cardiac Output Modules: M-COP and M-COPSv
•
NIBP Modules: M-NIBP, M-NESTPR, M-NETPR, M-NESTR, M-NE12STPR, M-NE12TPR and
M-NE12STR
•
Airway Modules: M-C, M-CO, M-COV, M-CAiOV, M-CAiO and M-CAiOVx
•
The Recorder Module, M-REC and the built-in recorders in F-CMREC or F-CMCREC may not
be used simultaneously.