Datex-Ohmeda AS/3 Compact Monitor, CS/3 Compact Monitor
48
Document No. 896 623
4.3
Interfacing Monitors via M-INT
It is possible to interface Datex-Ohmeda monitors, Critikon Dinamap 1846SX, Abbott Oximetrix 3,
Baxter Explorer and Vigilance, Nellcor N-100, N-200 and N-1000 to the Compact Monitor via the
Interface Module, M-INT. Interface Module, M-INT, has two serial and analog connectors (X7 and
X8). The parameters transferred are summarized in tables 4-1 and 4-2.
NOTE:
Using the Interface Module , M-INT disables interfacing with Datex-Ohmeda monitors via the
UPI/UPINET Board.
WARNING
Always make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
Table 2
Transference of Parameters, Other Monitors
Device
Waveforms
Numerics
Alarms
Critikon
Dinamap 1846SX
None
NIBP
None
Abbott
Oximetrix 3
None
SvO2/SaO2, CO
None
Baxter
Explorer
None
C.O., SvO2, REF, Tblood
None
Baxter
Vigilance
None
C.O., SvO2, C.C.O.,
Tblood
None
Nellcor
N-100
N-200
N-1000
Pleth (analog)
SpO2, Pulse rate
None
4.3.1
Connection to Datex-Ohmeda Monitors
Use the INT-External Device Cable.
1.
Make sure that the power to both devices is turned off.
2.
Connect the 9 pin D-shaped connector to one of the connectors on the Interface Module,
M-INT. Tighten the finger screws.
3.
Connect the 25 pin D-shaped connector to the corresponding connector on the other
monitor. Tighten the finger screws.