General Service Guide
55
Document No. 896 623
5
FUNCTIONAL CHECK
This instruction includes procedures for a functional check for Datex-Ohmeda AS/3 Compact
Monitor and CS/3 Compact Monitor running with the S-__99(A) software. However, most of the
procedures also apply to monitors with older monitor software.
The functional check is recommended to be performed after monitor installation.
Perform the procedures in the ascending order and skip the items that do not correspond to the
particular monitor.
The instructions include a check form (
Appendix A
) to be filled in when performing the
corresponding procedures.
The symbol
?
in the instructions means that the performed procedure should be signed in the
check form.
5.1
Recommended tools
Tool
Order No.
For product(s)
Sampling line 2.0 m
73318
Compact Airway Module, M-COVX/-
CAiOVX
SpO
2
finger probe
SAS-F4
Hemodynamic modules w/ (S)
Adult NIBP cuff
572435
Hemodynamic modules w/ (NIBP)
Adult NIBP hose
877235
Hemodynamic modules w/ (NIBP)
Nellcor SpO
2
probe
-
M-NSAT
Earphones
-
M-EEG
Tonometrics
TM
catheter
M-TONO
MemCard - Menu
-
M-MEM
MemCard - Data
-
M-MEM
Screwdriver
-
All
5.2
Visual inspection/preparation
•
Make sure that the monitor is turned off.
•
Disconnect the mains power cord and press the service reset- switch until the LED next to it
turns off.
•
If the monitor is connected to the Datex-Ohmeda network, disconnect the Mon-Net cable
from the monitor’s Network extension cable. If the Memory Board, B-CMNET or the Memory
Module, M-MEM, is connected, remove any memory cards.