![Datex-Ohmeda AS/3 Compact Monitor Technical Reference Manual Download Page 46](http://html.mh-extra.com/html/datex-ohmeda/as-3-compact-monitor/as-3-compact-monitor_technical-reference-manual_513828046.webp)
Datex-Ohmeda AS/3 Compact Monitor, CS/3 Compact Monitor
38
Document No. 896 623
3.
Unscrew the three screws holding the UPI Board in position and remove the board.
4.
Remove the UPINET Board, B-UPINET from the protective antistatic packaging. Always hold
the board by the edges and wear a wrist grounding strap.
CAUTION
The UPINET Board, B-UPINET comprises sensitive integrated circuits that can be
damaged by an electrostatic discharge. Careful handling of the board is therefore
essential.
5.
Insert the UPINET Board, B-UPINET into the empty slot and firmly press the board in
position. Secure the board with the two screws removed from the UPI Board earlier.
6.
Replace the back plate and secure the plate with the six screws removed from the plate
earlier.
7.
Connect the B-NET connector cable to the network connector.
8.
Reconnect the Power Cord.
NOTE:
These instructions applies only for F-CM(REC) rev. 04 and F-CMC(REC) rev. 01 Compact
Monitors. For earlier revisions also additional parts are required. Please contact your authorized
Datex-Ohmeda distributor.