68
Document No. 896 623
Invasive blood pressure measurement
55.
Check the module membrane
Zero P4
key.
Press the key for at least one second. Check that the key being pressed is identified, i.e. the
information on the service menu under ‘Button’ - ‘P4’ changes from
OFF
to
ON
.
?
SvO
2
measurement
56.
Check that the SvO
2
values ‘Meas. state’, ‘OM fail’ and ‘OM temp.’ in the
COP
module
service menu all show
NO OM
.
?
Cardiac Output measurement
57.
Check the
Start C.O.
module membrane key.
Enter the
COP
module service menu. Press the key for at least one second and check that it
is identified, i.e. the information in the service menu under ‘Button’ - ‘C.O.’ changes from
OFF
to
ON
.
?
NIBP module, M-NIBP
58.
Enter the NIBP module service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
-
Service
(password 26-23-8)
- Parameters - NIBP
Check that the Timeouts, Bad checksums and Bad c-s by mod values are not increasing
faster than by 50 per second. Check that the NIBP board memories have passed the
internal memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all show
OK
.
?
59.
Select
Pneumatics
.
Press the ComWheel twice. The NIBP pump should turn on and off.
?
60.
Check the module front panel LEDs and membrane keys using the
Buttons/Leds
service
menu:
Parameters - NIBP - Buttons/Leds
Highlight the
Auto ON
text. Check that the module front panel LED for the autocycle
measurement turns on and off when the ComWheel is pressed. Check the other LEDs by
selecting
Manual ON
,
Stat ON
and
Measur. ON
.