![Datex-Ohmeda AS/3 Compact Monitor Technical Reference Manual Download Page 65](http://html.mh-extra.com/html/datex-ohmeda/as-3-compact-monitor/as-3-compact-monitor_technical-reference-manual_513828065.webp)
General Service Guide
57
Document No. 896 623
Compact Airway Module, M-CXXXXX
5.
Disconnect the module and check external parts:
−
the front cover and the front panel stickers are intact
−
all connectors are intact and are attached properly
−
the D-fend latch moves properly
−
the module box is intact
−
the module latch is intact and moves properly
Plug the module back into the monitor.
?
Tonometry Module
6.
Disconnect the module and check the external parts:
−
the front cover and the front panel sticker are intact
−
connectors are intact and are attached properly
−
the module box, latch and spring locking pin are intact
Plug the module back into the monitor.
?
Other plug-in modules and N-EEG
7.
Unplug the modules one by one and check external parts:
−
the front cover and the front panel sticker are intact
−
all connectors are intact
−
all mechanical parts are functioning properly
−
the module box and the latch are intact
−
the module bus connector is clean and intact
Plug the modules back into the monitor.
?
8.
Check external parts of the EEG Headbox, N-EEG and connect the EEG Headbox to the EEG
Module.
?