64
W-Class
Advanced Entry Calibration (continued)
Step
Operator Action
Displayed Message
Comment
16
Press the
key.
Using the or the
key, set the
‘Paper’ level to the value
determined in the previous
step. Press the
key
to set the entry (indicated
by the ‘*’) and advance the
menu.
Repeat for the ‘Gap’ (or
‘Mark’) and ‘Empty’ levels.
PAPER SENSOR LEVEL
P*198 G*084 E*014
Ç
GAP SENSOR LEVEL
P*198 G*084 E*014
Ç
EMPTY SENSOR LEVEL
P*198 G*084 E*014
Or, for reflective media:
PAPER SENSOR LEVEL
P*015 M*181 E*213
Ç
MARK SENSOR LEVEL
P*015 M*181 E*213
Ç
EMPTY SENSOR LEVEL
P*015 M*181 E*213
The selection
will flash and
the display
heading will
change to
indicate the
item for entry.
17
After all entries have been
made, press the
key to back out of the menu
and then press the
key
to save the settings and
return to the Ready Mode.
SAVE CHANGES?
ENTER KEY = YES
From ‘Ready’,
press the FEED
key to advance
to the next label
TOF.
ã
Note: If the Advanced Entry Calibration fails, try the following procedure.
Re-enter Media Settings / Calibration / Advanced Entry / Sensor Gain
and lower the selected Gain Setting number by one (to make the
sensor less sensitive). Select the new Gain Setting, save the changes,
exit the menu, and then test your media. Repeat the procedure until a
usable media setting is obtained.