56
W-Class
Standard Calibration (continued)
Step
Operator Action
Displayed Message
Comment
3
Press the
key to
enter the Media Settings
menu.
MEDIA SETTINGS
MEDIA TYPE
See Section 4.1.2
for menu layout.
4
Press the key
and scroll to ‘Sensor
Calibration’.
MEDIA SETTINGS
SENSOR CALIBRATION
Press the
key to abort this
procedure.
5
Press the
key.
SENSOR CALIBRATION
PERFORM CALIBRATION
You are beginning
the standard
calibration
procedure.
6
Press the
key.
Remove the media from
the Media Sensor then
press any key.
REMOVE LABEL STOCK
PRESS ANY KEY <yyy>
This sets the
parameter for the
‘empty’ value.
Where ‘yyy’ is a
numerical value
representing the
current sensor
reading.