A-10
POWERSCAN D8590 User’s Manual
Troubleshooting
Problem: No Illumination – License Key Failed.
The unit is unable to decode, and it is not cycling through its illumination settings. (Targeting,
if enabled, does appear when the trigger is pulled.)
Error Messages (String Output):
“License Key Failed”
Possible Cause:
The POWERSCAN D8590 license has been deleted.
Solution:
Request license file (2019_(serial number).crb) from Datalogic customer support.
Unit serial number and license must match.
Once you have received the license file
from customer support, go to the
Utilities
section of ESP, browse for the file in the
Batch File Update
field, and click
Start
once you have located the file.
Problem: No Illumination – No Process Sequence Defined.
The unit is unable to decode, and it is not cycling through its illumination settings. (Targeting,
if enabled, does appear when the trigger is pulled.)
Error Messages (String Output):
“No Process Sequence Defined; Please Set the Unit to
Default Settings”
Possible Cause:
No illumination sequence has been defined for the POWERSCAN D8590.
Solution:
Reset the default illumination sequence.
If
Sequence Steps
is set to
0
, the illumination
sequence will not occur.
Click the
Default
button to
return to
4
illumination steps.
Summary of Contents for PowerScan PD8590-DPM
Page 1: ...PowerScan PD8590 DPM Imager Industrial Corded Handheld Bar Code Reader User s Manual...
Page 32: ...2 18 POWERSCAN D8590 User s Manual Send Receive NOTES...
Page 42: ...3 10 POWERSCAN D8590 User s Manual Illumination Sequence NOTES...
Page 58: ...4 16 POWERSCAN D8590 User s Manual Other Communications Settings in ESP NOTES...
Page 64: ...5 6 POWERSCAN D8590 User s Manual Region of Interest NOTES...
Page 84: ...6 20 POWERSCAN D8590 User s Manual Symbology Identifier NOTES...
Page 110: ...9 8 POWERSCAN D8590 User s Manual Terminal Dropdown Menu NOTES...
Page 121: ...POWERSCAN D8590 User s Manual A 5 Appendices USB Cable Pinouts 1 828 80 mm...
Page 127: ......