ENC58 PROG
29
5.5 Error messages
(in alphabetical order)
Activation error of sensor data!!!!!
The operator has pressed the READ button but an error has
occurred while processing the information (it is signalled through the
LED lit red in the diagnostic information section of the page).
Please close the programming interface, then turn off and on the
power supply of the encoder (disconnect and connect again the USB
cable), finally restart the programming interface.
Beware!!! Displayed parameters could differ from encoder
parameters!!
The operator has pressed the Load config button (or otherwise as an
alternative the Load config command in the File menu) in order to
upload a previously saved encoder configuration.
After having select and confirmed the .cfg or .hex file to upload, the
Would you like to update also encoder parameters?
message
appears on the screen after a few seconds.
If you choose YES, the configuration parameters are updated in
both the programming interface and the encoder; otherwise, if you
press NO, the configuration parameters are updated only in the
programming interface, they are not downloaded to the encoder.
In the
Messages
section below in the page the
Beware!!! Displayed
parameters could differ from encoder parameters!!
message warns of
occurrence and the Load config button starts blinking red.
Beware!!! Voltage output will be changed!!! Continue?
The operator is going to change the voltage level at the output
channels by programming the
Voltage output
parameter.
This message is intended to ask the operator to confirm before
proceeding.
Device not connected!!!!! Click on 'Disconnected' button
If you try to send a command while the encoder is not connected yet,
then the
Device not connected!!!!! Click on 'Disconnected' button
warning message will appear on the screen.
Press the DISCONNECTED button to activate the communication
and start programming the encoder.
Encoder disconnected!!! Check encoder connections
This warning message is triggered when the interface establishes a
proper USB connection, but it is not able to communicate with the
encoder.
When this happens, please check either the M23 connector.