Water Flow Meter Calibration:
A. Ensure tube is filled with water.
Go to:
MANUAL MODE, MANUAL FUNCTION, FILL CW 50ml
to fill tube with
water.
B. Drain beaker.
Go to:
MANUAL MODE, MANUAL FUNCTION, DRAIN XXsec.
Ensure
the beaker lid is on to build pressure up so water is pushed out of the
beaker.
C. Release pressure in the chamber by pressing the manual “Pressure Relief” valve
switch located next to the chamber as shown below.
Note: It is important to maintain atmospheric pressure during calibration for
accuracy.
ahiba
49
AHIBA MULTIPRECISE TC - Calibration (Water Flow Meter)
Summary of Contents for Ahiba Ahiba MULTIPRECISE TC
Page 1: ......
Page 3: ......
Page 6: ...ahiba 5 Safety Precautions Safety Precautions and W and Warnings arnings ...
Page 7: ......
Page 11: ......
Page 12: ...ahiba 11 Introduction T Introduction Typical ypical Applications Applications ...
Page 13: ......
Page 16: ...ahiba 15 Main Components Main Components ...
Page 17: ......
Page 25: ......
Page 26: ...ahiba 25 Installation and S Installation and Start Up tart Up ...
Page 27: ......
Page 35: ......
Page 36: ...ahiba 35 T Technical Data echnical Data ...
Page 37: ......
Page 41: ......
Page 42: ...ahiba 41 Procedure for Dosing Procedure for Dosing ...
Page 43: ......
Page 48: ...ahiba 47 Calibration Calibration ...
Page 49: ......
Page 53: ......
Page 54: ...ahiba 53 Controller Controller ...
Page 55: ......
Page 81: ......
Page 82: ...ahiba 81 Protection Protection ...
Page 83: ......
Page 85: ......
Page 86: ...ahiba 85 Maintenance Maintenance ...
Page 87: ......
Page 93: ......
Page 94: ...ahiba 93 Appendices Appendices ...
Page 95: ......