3 … Troubleshooting
Fiducial Scan Errors
—29—
Figure 16—PNP Head optic sensor digital display
showing a reading of 12.
The reading on the
top surface of the module
should be approximately in the range
1000--2500. If it is too high (>2500), then roughen
the top surface with 400 grit sandpaper. If the
reading is too low ( <1000), then polish the top
surface with fine emery cloth.
If the reading on the
fiducial slot
is not within
the allowed range (0– 500), gently blow out the
fiducial slot with clean, dry, compressed air.
NOTE: When running the Fiducial Scan
from the Diagnostic Window, repeatability
on subsequent scans should be ±10 counts.
(One count is 1/100 mm.)
If problems with the fiducial scan continue, see
the on-screen Diagnostics Help, which includes
further diagnostics. You need a PIN to enter the
Diagnostics Windows (Tools button) prior to tap-
ping the Help button.
Summary of Contents for FLX500
Page 4: ... Data I O FLX500 ii ...
Page 12: ... Data I O FLX500 Chptr 1 8 blank page ...
Page 28: ... Data I O FLX500 Chptr 2 24 blank page ...
Page 40: ... Data I O FLX500 Chptr 4 36 blank page ...
Page 42: ... Data I O FLX500 Chptr 5 38 Daily Sign off Clean the sockets page 10 Date Initials Notes ...
Page 43: ...5 Sign off Charts Daily Sign off 39 Daily Sign off Date Initials Notes ...
Page 44: ... Data I O FLX500 Chptr 5 40 Daily Sign off Date Initials Notes ...
Page 45: ...5 Sign off Charts Daily Sign off 41 Daily Sign off Date Initials Notes ...
Page 46: ... Data I O FLX500 Chptr 5 42 Daily Sign off Date Initials Notes ...
Page 48: ... Data I O FLX500 Chptr 5 44 Weekly Sign off Date Initials Notes ...
Page 52: ... Data I O FLX500 Chptr 5 48 blank page ...
Page 57: ... FLX500 ...