Copyright © 2008 – 2019 DARE!! International
Page 32
Wait for the “OK” response.
Verify that the table is programmed correctly into the RadiSense 10 probe by dumping the table.
Send the “CALDUMP” command. The response should be the table which is programmed by the
“CALSTORE” command.
Turn the correction factors ON. This can be done by using the “CAL ON” command.
The response should be: “OK”