5.3
Error codes - user correction table
During uploading of the correction file the following error messages might be returned.
Error message
Cause
Incorrect column amount
A wrong number of columns is used in the correction file.
Incorrect precision
A value in the correction file has a too high precision.
Correction factor out of range A value in the correction file is greater than 3.00.
Wrong columns available
A wrong column is used in the correction file.
Incorrect number of rows
Too much frequency points are used in the correction file.
Contains negative values
A value in the correction file is negative.
Frequencies are incorrectly
sorted
One or more of the frequencies are sorted from high to low.
Multiple same frequencies
Multiple of the same frequencies are used in the correction
file
Correction file is not found
The selected correction file could not be found.
CAL Dump is not matching
The uploaded correction file does not match the returned
table from the RadiSense 10
Correction file could not be
opened
The selected correction file could not be opened.
Correction file is empty
The selected correction file is empty.
Invalid low frequency
The selected correction file contains a too low frequency.
Frequencies within rounding
distance
Multiple frequencies shall be rounded to the same frequency
Invalid upper frequency
The selected correction file contains a too high frequency.
RadiSense 10 is an old firmware
version
The firmware of the connected RadiSense 10 is less than
“2.0.6”
RadiSense 10 CAL dump is not
returned
The RadiSense 10 failed to return the uploaded calibration
table back.
RadiSense 10 CAL table
is invalid
The RadiSense 10 could not validate the stored calibration
table.
Please refer to the RadiCentre manual for an overview of all error codes.