Copyright © 2008 – 2019 DARE!! International
Page 43
RadiSense Specifications, part 2
Model
RadiSense
10
Optical
LASER power
Max. 0,5 Watt output at aperture
Wavelength
808 nm
LASER connector
FC/PC
Data connector
ST/PC
Fibres
200/230 µm HCS, duplex
Standard Fibre length
1,5 m fixed to sensor, 10 m extension with couplings
Other lengths available on request
Safety
LASER product
classification
Class 1M according to EN 60825-1 and EN60825-2
Safety measures
Code (RadiCentre),
Remote interlock system (RadiCentre),
LED indications for LASER ON,
Audible warning signals
Redundant closed loop safety system
LASER switch on time
approx. 50 ms
LASER switch off time
< 10 ms
Environmental conditions
Temperature range
0° to 40° Celsius
Relative humidity
10 – 90% (non-condensing)
Warranty
Warranty
3 years (misuse excluded)