8.5   Pulse input function module

Module for two additional meters. Pulse input 1 is marked as “I1-   ”, 
input 2 with “I2 -   “. Pulse inputs can be programmed (IZAR@SET) with 
a value: 1, 2.5, 10, 25, 100, 250, 1000, 2500 litre per pulse.
•  Pulse transmitter must be electrically isolated, e.g. Reed contact
•  Possible units are all the energy units available in the meter, the 

volume unit m³ or no unit.

Data is accumulated separately in registers; can be read in the dis-
play  as IN1 and IN2 and can be transferred via the communication 
modules.

8.6   Pulse output function module

The module is equipped with 2 pulse outputs, which can be freely 
programmed using the IZAR@SET software. The outputs are marked 
as “O1 -   “ and “O2 -   “ on the terminal strip and as Out1 and Out2 
on the display.
•  External supply: Vcc = 3-30 VDC
•  Output current ≤ 20 mA with a residual voltage of ≤ 0.5 V
•  Open Collector (Drain)
•  Electrically isolated
•  Output 1: f ≤ 4 Hz Pulse duration: 125 ms ±10 % 

Pulse pause: ≥ 125 ms –10 %

•  Output 2: f ≤ 100 Hz Pulse duration pulse/pause ~1:1
•  The volume pulse rate can be freely programmed
•  Standard: last digit in the display

8.7   Combined function module (IN/OUT)

The combined module has 2 inputs and 1 output.
See chapter 8.5 for the specific characteristics on the pulse input. 
A 3 VDC voltage is connected to the “+” terminal and can be used as
the supply for a flow sensor.
The pulse output has the same characteristics as pulse output 1 speci-
fied under 8.6. It is not, however, electrically isolated.

8.8   Analogue output function module

The module has 2 passive analogue output connections which are 
freely programmable using the IZAR@SET software. On the terminal 
strip, the electrically isolated outputs “1” and “2” are marked and the 
polarity is indicated (“+” and “–”).
•  Passive, external power supply: 10…30 VDC
•  Current loop 4 … 20 mA whereat 4 mA = 0 value; 

20 mA = programmable maximum value

•  Overload up to 20.5 mA, then residual current
•  The module displays errors with 3.5 mA or 22.6 mA 

(programmable)

•  Output values: power, flow, temperatures

8.9   Test output

The internally located test output is intended for testing 
laboratories. The manufacturer supplies two special cables:
1. Volume testing pulse
2. Energy testing pulse
Please refer to the inspection and testing manual for further 
specifications (pulse rate, pulse duration/pause, pulse frequency).

Input frequency

≤ 8 Hz

Min. pulse duration

10 ms

Input resistance

2.2 MΩ

Terminal voltage

3 VDC

Cable length

up to 10 m

                                

                                

                                              

                                      

                                       

The module is connected with the meter electronics by means of
a ribbon cable. The separate plug connector on module slot 2 is
necessary for the perfect functioning of the analogue outputs.

Ensure that the temperature sensors (measurement resistances)
remain in contact without interruption during energy verification.

T

T

T

T

Danfoss District Energy 

 

 

 

 

VI.SH.K1.02 

 

 

 

               DEN-SMT/PL

7

Summary of Contents for SONOMETER 1100

Page 1: ...Sheet FW510 If water additives are used e g corrosion protection the user must make sure that the corrosion resistance is adequate The specified medium temperature is 5 90 C 105 C 130 C 150 C The tem...

Page 2: ...low sensor in a tilted 45 position The minimum system pressure must be 1 bar to avoid cavitation Make sure the energy meter is installed sufficiently far away from possible sources of electromagnetic...

Page 3: ...be inserted as far as the bottom of the pocket and then secured The pockets are best installed inT pieces with a 45 or 90 angle The tip of the pocket must point in the opposite direction to the direct...

Page 4: ...on pressing button and the date and time are still updated but none of the measuring functions work incl the flow rate measurement Communication still functions over the optional M Bus RS485 and RS23...

Page 5: ...e formed ribbon cable at both ends 3 Close the lid and check the meter for correct operation by pressing the push button loop 3 Renew the seal of the housing lid if the meter functions correctly The r...

Page 6: ...tandard and is encrypted Reading modes Walk By Drive By Fixed Network For problematic radio installations shield the external radio module set can also be used 8 2 M Bus communication module The M Bus...

Page 7: ...nd can be used as the supply for a flow sensor The pulse output has the same characteristics as pulse output 1 speci fied under 8 6 It is not however electrically isolated 8 8 Analogue output function...

Page 8: ...Future accounting date 1 2 3 Accounting date 1 previous year Accounting date 1 previous year energy Accd 1L 2 4 Accounting date 2 Accounting date 2 energy Accd 2A 2 5 Accd 2 Future accounting date 2...

Page 9: ...prox 4 minutes The error display disappears automatically as soon as the cause of the error has been cleared All errors present longer than 6 minutes are saved in the error log Error code Description...