Piranha HS-S0-06K Camera User Manual
Teledyne DALSA
03-032-20213-01
55
Mnemonic
Syntax Parameters Description
set d ata w id th
sdw
i
Selects the bit d ep th, w here:
sd w 8 = 8-bit ou tp u t
sd w 10 = 10-bit ou tp u t
sd w 12 = 12-bit ou tp u t
set exp osu re m od e
sem
m
Set the exp osu re m od e:
3 =
External SYN C, m axim u m
exp osu re tim e
7 =
Internal p rogram m able SYN C,
m axim u m exp osu re tim e. Factory
setting.
set fp n coeff
sfc
x i
Set the FPN coefficient.
x
=p ixel nu m ber w ithin the range
1
to
6000
.
i
= FPN valu e w ithin the range
0
to
8191
(12-bit LSB).
Refer to Perform ing FPN Correction on
p age 30 for d etails.
set gain
sg
tf
0 to 30 : -20 to +20.
set p rnu coeff
spc
x i
Set the PRN U coefficient.
x
=p ixel nu m ber w ithin the range
1
to
6000
.
i
= PRN U valu e w ithin the range
0
to
65535
set su btract backgrou nd
ssb
i
Su btract the inp u t valu e from the
ou tp u t signal.
i
= Su btracted valu e in a range from
0
to
4096
set sync frequ ency
ssf
i
Set the fram e rate to a valu e from :
TDI
1 to 142,000.
Area
1 to 320.
Valu e rou nd ed u p / d ow n as requ ired .
Refer to Setting Fram e Rate on p age 25
for d etails.
set system gain
ssg
i
Set the d igital gain.
i
= Digital gain in a range from
0
to
61438
. The d igital vid eo valu es are
m u ltip lied by this nu m ber. Refer to
Setting System Gain on p age 34 for
d etails.
set set nu m ber
ssn
i
0 to 4.
stage select
stg
i
Sets the nu m ber of TDI stages.
For the H S-S0, the available valu es are
64, 128, 192, 256.
Factory setting
is
256
. Refer to Selecting the N u m ber of
CCD Integration Stages for d etails.