ERROR MESSAGES
MESSAGE
PROBABLE CAUSE
TROUBLESHOOT
No error
-
-
Flash memory integrity error
Troubles occurred with the non-volatile
memory
U•V6 control unit maintenance
Display not ready
Troubles occurred with the display board
U•V6 control unit maintenance
Display initialization error
Display error
Keys & leds not ready
Keys & leds initialization error
EEPROM not ready
An operation with the EEPROM failed
U•V6 control unit maintenance
EEPROM reading error
EEPROM integrity error
EEPROM writing error
Date saving error
Lowest temperature saving error
Highest temperature saving error
Temperatures saving error
Password loading error
Password saving error
Password update error
Date & time setting error
The operation failed
Retry. If the error is persistent,
U•V6
control unit maintenance
System configuration initialization error
Something went wrong during initialization
Reboot. If the error is persistent, U•V6
control unit maintenance
Core configuration load error
History initialization error
Date initialization error
ADC calibration error
Data acquisition failed
U•V6 control unit maintenance
ADC conversion (internal) error
ADC conversion (config.) error
ADC conversion (DMA) error
ADC conversion timeout
Power failure
The U•V6 control unit is not correctly
supplied by the main power
Check the the main power supply
Battery missing
The
U•V6 control unit does not view the
battery
U•V6 control unit maintenance
Battery low voltage
The battery need to be charged. The main
power supply is expected
If the error is persistent, U•V6 control unit
maintenance
Battery discharged
The battery need to be charged. The main
power supply is expected. The control unit
shall halt
If the
error is persistent, U•V6 control unit
maintenance
Defective battery
The battery cannot be managed by the
control unit
U•V6 control unit maintenance