U•V6
Instruction Manual
15
|
65
Certified
ISO9001 • VCA
See the operation manual for the siren for instructions on how to connect the siren and
configure its tone.
2.7
Connecting the Analogue Output
Connect the
S
and
V
–
terminals on the analogue output to close the circuit. In current-loop
mode, the current leaves via terminal
S
and returns via terminal
V-
.
The analogue output is configured as follows before leaving the factory:
▪
In 4-20 mA current-loop mode
▪
The output signal corresponds to the maximum value of the 6 input channels
▪
The output gain ratio is 1:1 of the detectors' measurement range. A full-scale
measurement of the detector is shown as 20 mA on the analogue output signal
The user can change the unit from a current-loop signal mode to a voltage-signal mode by
toggling the
RL Enable
switch:
▪
The
ON
position of the
RL Enable
switch transforms the output signal to a voltage
signal, by adding a 500Ω shunt resistor.