PVX-B
Precision Visual Thickness Gauge
33
One Point Calibration
Note:
Be sure that a probe zero has been done prior to performing this
calibration procedure if a contact transducer in pulse-echo mode is being used.
1)
Physically measure
a location on the sample to be measured using a set of
calipers or a digital micrometer. Note: It’s always best to calibrate on
thickest location on the sample, rather than on the thinnest location, as the
percentage error is much less with respect to the
PVX-B’s
resolution of +/-
.0001” (0.001mm).
2) Apply a drop of couplant on the transducer. Place the transducer in steady
contact on the known thickness location. Be sure that the reading is stable
and the repeatability indicator, located in the top left corner of the display, is
fully lit and stable. Press the
MENU
key once to activate the menu items tab.
Press the
MENU
key multiple times to tab right and the
ESC
key multiple
times to tab left until the
CAL
menu is highlighted and displaying the
submenu items.
3) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
ONE POINT
is highlighted.