7
. Output Setting Guide37
7. Output Setting Guide
Part
Cut Mode
Cut Power
Coag. Mode
Coag. Power
BI-Coag.
Skin Incision
Pure
or Blend1
10
~
120
8
~
100
Muscle
Dissection
Pure
or Blend1
Above 15
Above 15
Tumor
Excision
Blend3
Above 10
Stomach,
Intestine
Resection
Blend2
or Blend3
20 and up
20 and up
Hemostasis
Contact
10
~
65
3
~
22
Massive
Coagulation
Spray
10
~
30
Neuro Surgery
Blend2
or Blend3
Loop
20
~
80
20
~
70
Contact
BALL
10
~
25
1
~
8
Prostatic
Resection
Pure
or Blend1
65 and up
55 and up
Spray
15 and up
Bladder
Fulguration
Spray
12
~
30
Tumor Excision
Blend2
Blend3
15
~
80
15
~
70
Spray
10
~
15
Cervical
Conization
Blend2
10
~
80
Contact
25
~
70
Bartholin
and Skeneis
Blend3
15
~
30
Spray
12
~
22
Tubal
Blend2
8
~
50
10
~
20
Proctologic
Blend3
8
~
40
Spray
10
~
22
Abscess/Cyst
Blend3
10
~
80
Spray
10
~
15
Rectal, Sigmoid Blend3
Lancet 8
~
30
Loop 10
~
20
Contact
12
~
30
Endoscopic
surgery
Pure
25-40
Spray
30-50
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...