3. Specifications
3.1 Specifications
Product Name
Electrosurgical Equipment
Model
DT-400S/300S
Purpose
Cutting or coagulation of biological tissue using high-frequency current
Mains and Line
Frequency
120(60Hz) or 230VAC(±10%) (50 or 60Hz)/switchable
See the label on rear side for correct information
Fuses
Two of F5.0A, 250V for 230VAC
Two of F10.0A, 250V for 115 VAC
Power Consumption
900VA
Protection Class
Class 1, Type CF
IPX
Foot Switch: IPX6
Carrier Frequency
400 KHz ±10%, 482KHz ±10%
Repeat Frequency
33KHz
Low Freq. Leakage
Current
EN60601-1 [2012] Requirement
High Freq. Leakage
Current
Less than 150mA
Operation Conditions
10
℃~
40
℃
, 30%
~
85
%
noncondensing, 700mbar
~
1060mbar
Storage Conditions
-10
℃
~
60
℃
, 30%
~
95
%
noncondensing, 700mbar
~
1060mbar
Transportation
Conditions
-10
℃
~
60
℃
, 30%
~
95
%
noncondensing, 700mbar
~
1060mbar
Do not drop the unit from higher than 50cm
Do not use the hook.
Installation
Conditions
5cm clearance required on each side unit for cooling
Cooling
1 fan mounted
Recommended Duty
Cycle
10 sec ON, 30 sec, OFF(Output for 10 seconds, pause for 30 seconds )
Size (H, W, D)
147mmx 330mm x 417mm
Weight
About 16Kg
APG
Not AP/APG device
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with or with oxygen or nitro oxide
Operation Sound
Difference in operation sounds between cutting and coagulation.
Operation Display
Visible difference in operations between cutting and coagulation.
Output Setting
Displayed in [W] unit from Seven Segment (LED number displayer)
※
It can be
changed without notice
to improve performance.
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...