6
. Operating and Sequence31
3) Select Single Pad Mode or Dual Pad Mode
▶
Single Pad Mode
①
Connect the Single Pad to the Return Electrode Jack. The Alarm will stop and it will be
ready to be used.
②
Attach the Pad firmly to the patient. (Please read carefully the precautions)
▶
Dual Pad Mode
①
Connect the Dual Pad to the Return Electrode Jack.
②
Press
(Single Pad or Dual Pad Selection Switch) to set up the Dual Pad Mode.
The Dual Pad Mode Indicator will be turned on and the DRM will be ready.
③
Attach the Pad firmly to the patient. (Please read carefully the precautions)
④
The changes in the Assist Resistance Indicatorwill be seen. You can set DRM by pressing
(DRM Setting Switch) if the LED light on from 3 to 8 steps.
※
The Assist Resistance Indicator has 2 steps or 0 step when the Return Electrode cable
may be cut off or the Pad is not attached properly to the patient. The Indicator has 8 steps
when the Return Electrode cable may be short-circuited.
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...