10
. User Maintenance54
- Checking whether there is damage to Bi-polar Forceps, TBH, Unipolar Handle, and Neutral
Plate including the checkup the damage of any accessories and cleaning status.
- Obvious external damage to the unit.
(3) This equipment requires the inspection on the following items at least once a year.
- Obvious internal damage to the unit.
- Accumulation of the lint or debris within the unit or heat sink.
10.5 Periodic Performance Testing
This equipment requires the performance test at least once a year.
- Check if the output is accurately generated. When inaccurate output is detected, please
contact sales agents or DAIWHA Corporation for getting the calibration service.
- Check if the ANPM function is accurately operating.
TIP! Please, refer to the What is ‘
Tip: WHAT iS D.R.M (Daiwha Return electrode
Monitor)?
’section?
10.6 Product Life Span
Product life span may vary according to usage practice and usage conditions.
- All disposable accessories: One time use only.
- All reusable accessories : 50 times use only
- Recommended product life span: 2 years
* If the unit package indicates maximum years of the use, follow that information.
* Check all reusable parts with eyes before use. Scrap immediately or repair them before use.
As for the reusable parts, the unit life span is significantly different according to usage
environment and sterilization method, there is not warranty period allocated.
* If 5 year has been passed since usage, contact DAIWHA Corporation, LTD. for getting the
inspection service to determine whether to continue operation. Also, do not miss the
periodic inspection on the safety and performance.
10.7 How to Scrap the Unit and Accessories
- Discard the ESU when the time limit is passed.
- Scrap infected parts safely not to cause second infection, following the rules and regulations in
the corresponding laws.
10.8 Warranty
The manufacturer warrants that the ESU is free from defects in material and workmanship
under normal use and service for a period of months after the purchase data.
The manufacturer its authorized representative takes obligation to carry out the warranty
repair of the ESU or to replace the ESU with an operational one in case the Manufacturer
or its authorized representative determines that the cause of the ESU's failure was related
to the manufacturing process.
If the Buyer finds the defect in the ESU during the warranty period, he must report it and
inform the Manufacturer or its authorized representative within 30 days.
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...