6
. Operating and Sequence36
4) Scrap disposable accessories in the prescribed regulations (according to the law). Exert your
caution in a way that there is no risk of contamination due to the used consumable accessories.
6.3.4 Precautions during use
1) Precautions during use
- Check if the current output setting is appropriate to the existing surgical environment including
the surgical area.
- Make sure that patients do not touch grounded metals. It may cause partial burn.
- Do not give long time output to tissue areas or higher output beyond the need. Manipulate
output when electrode is always contacted with tissue to prevent tissue burn.
- If the patient complains about electric shock, stop using the unit immediately and have it
examined thoroughly.
- Frequently check contact status between neutral plate and a patient.
- The sound volume should be at least the medium of level to prevent sound-recognition-fail
caused dangers.
2) Storage condition or storage method
- Storage Condition: -10
℃
~
60
℃
, 30%
~
95
%
noncondensing, 700mbar
~
1060mbar
- Put in the product package box, and place in the cool area.
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...