3
. Specifications22
4) Alarm
: If Return Electrode is dismounted from the device or its cable is broken, Return Fault
Indicator flickers and alarm tone sounds, cutting off the output.
- For dual pad
1) Function
: It monitors the contact area of pad as well as mounting and cable breaking status of
Return Electrode.
2) Component
: DRM Setting Lamp, DRM Setting Switch, Single pad or Dual pad Selection Switch,
Dual pad Mode Lamp, Assist Resistance Indicator, Return Fault Indicator
3) How to use
①
'Single pad Mode Lamp' lights when applying power to the device as the default is
single pad mode. Press
(Single pad or Dual pad Selection Switch) in order
to switch on ‘Dual pad Mode Lamp’.
②
Mount Return Electrode to Return Electrode jack.
③
Patch the pad on the patient's body as closely as possible so that more Assist
Resistance Indicators could light for the safety, if possible. (Refer to the ‘5) Relation
between the light counts of Assist Resistance Indicator and pad resistance’ in the next
page.)
④
See ‘Assist Resistance Indicator’ change. If more than three lights are turned on, it is
possible to use them.
4) Alarm
(1) Error
-
I
f
a
n
error occurs, check out the error number displayed in equipment, and contact DRE Isales
agent or DRE
(2) Alarm
Alarm
No.
Alarm
Display
Alarm Situation
How to cancel the alarm
1
“ALr 1"
When connecting ‘Dual Pad’ while using
the single pad mode
Connect ‘Single Pad’ to ‘Neutral
plate’ jack.
When the ‘Single Pad’ is generated from
‘Neutral plate’ jack while using the single
pad mode
In case of error occurrence
In case of alarm occurrence
Summary of Contents for Meditom DT-400S
Page 29: ...5 Connecting Accessories with the unit29 5 3 Bipolar Forceps 5 4 Foot switches ...
Page 40: ...9 Load Regulation Vpeak Graphs40 9 1 2Open Circuit Peak Voltage ...
Page 41: ...9 Load Regulation Vpeak Graphs41 9 2 Blend1 Cutting Mode 9 2 1Load Regulation DT 400S DT 300S ...
Page 42: ...9 Load Regulation Vpeak Graphs42 9 2 2Open Circuit Peak Voltage ...
Page 43: ...9 Load Regulation Vpeak Graphs43 9 3 Blend2 Cutting Mode 9 3 1Load Regulation DT 400S DT 300S ...
Page 44: ...9 Load Regulation Vpeak Graphs44 9 3 2Open Circuit Peak Voltage ...
Page 45: ...9 Load Regulation Vpeak Graphs45 9 4 Blend3 Cutting Mode 9 4 1Load Regulation DT 400S DT 300S ...
Page 46: ...9 Load Regulation Vpeak Graphs46 9 4 2Open Circuit Peak Voltage ...
Page 48: ...9 Load Regulation Vpeak Graphs48 9 5 2Open Circuit Peak Voltage ...