![Daiwha DI-6000 User Manual Download Page 72](http://html1.mh-extra.com/html/daiwha/di-6000/di-6000_user-manual_3237401072.webp)
Appendix
66
10.
Appendix
10.1
EMC declaration
Guidance and Manufacturer's Declaration - electromagnetic emissions
Infusion Pump DI-6000 is intended for use in the electromagnetic
environment specified below.
The customer or user of the pump should assure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF Emissions
Group 1
Infusion Pump DI-6000 uses RF energy only for their int
ernal function. Therefore, their RF emissions are very lo
w and are not likely to cause any interference in nearby
electronic equipment.
CISPR 11
RF emissions
Class A
Infusion pump DI-6000 is suitable for use in all establis
hments, including domestic establishments and those dir
ectly connected to the public low-voltage network that s
upplies buildings used for domestic purposes.
CISPR 11
Harmonic Emissions
Compliance
IEC 61000-3-2
Voltage Fluctuations
/
Flicker Emissions
Compliance
IEC 61000-3-3