Technical
Specification
61
9.2
Functional specifications
Function
Description
Alarm
Air bubble alarm
Down-Steam Occlusion alarm
Up-Stream occlusion alarm
Door Open Alarm
Low battery alarm, empty battery alarm
AC/DC removal power alarm
Standby alarm (interval of 2 minutes, with no injection in progress)
Near complete alarm
Injection is stopped upon alarms except Low Battery and Start Reminder.
Drop alarms (No Drop and Free Drop)
Malfunction
– Product state indicators are turned on and alarms are generated.
Safety
controls
Door Open:
Injection is disabled while the door is open.
Key Lock Button:
The buttons except
START/STOP, POWER,
and
LOCK
buttons do not work.
UP/DOWN occlusion sensor
: Detects occlusion in the I.V Set.
Nurse Call:
Allows monitoring upon alarms.
Air Sensor:
Detects air bubbles.
Drop Sensor:
Detects infusion drops.
Others
Time Mode:
Calculates the flow rate based on the injection time settings.
Dosage Mode:
Calculates the flow rate based on the dosage settings.
Titration:
Allows changing the flow rate while in injection.
Secondary Injection Mode (Primary/Secondary)
Low, Empty Battery:
When running on battery, the alarm occurs when the
remaining battery time is down to 30 minutes and 3 minutes.
Date and time:
The date and time of the dedicated clock are displayed with
no power supply.
Standby mode:
Pause function (1 min-24 hours, increment of 1 minute)
Event History:
Up to 2,000 events
Alarm Log Event:
Up to 50 alarm logs
Electronic data on the infusion pump is preserved for more
than 2 years with power off.