![Daiwha DI-6000 User Manual Download Page 37](http://html1.mh-extra.com/html/daiwha/di-6000/di-6000_user-manual_3237401037.webp)
Operation
31
5.2
Standby status (before operation)
In this status, if you press the
START
button in standby mode, operation is immediately
started.
With I.V Set being loaded, the pump can be changed to the standby mode by pressing the
POWER button for 3 seconds.
You can enter any status from this status.
When the Key LOCK switch is set to On, see
5.6 Key lock button function.
If no operation setting (button operations) within the time that has been set in standby
mode, a standby alarm rings and the red
LED
blinks. (See
7.14 Standby alarm
)
If you press the
BOLUS
button, injection is performed with the preset BLOUS flow rate and
volume.
Caution
If you press the
BOLUS
button in standby mode, it detects the occlusion as a
purged status.
Be careful because overdose can happen momentarily when using the
BOLUS
Button.
5.2.1
Setting the I.V. Set Code
Set the code name of the I.V Set before using it.
In the I.V Set menu, enter password to set the code
name. (PASSWORD : 1711)
The code name you set will be stored in internal
memory for future use.
Make sure the I.V Set matches the code name and it is
properly loaded before using.
See
3.3 Mounting I.V Set
Summary of Contents for DI-6000
Page 1: ......
Page 44: ...Custom Settings 38...
Page 71: ...Technical Specification 65...
Page 77: ......
Page 78: ......