Installation
16
3.3
Mounting I.V Set
1. Place the roller clamp of the I.V Set at the
lower side of the product and lock it.
Insert the guide needle into the I.V bag and fill the
chamber of the set with I.V liquid up to 1/3
or 1/2.
Open the roller clamp to remove all air within the
tube and lock the roller clamp.
2. Mount the I.V container on the pole.
At this time, do not place the I.V container at a
place 1 m or higher from the patient's heart.
3. Pull the door clamp to open the front part.
4. Grab the detector for Upstream Occlusion
by hand and firmly press it to fix the tube.
5. Press the I.V Set to fix the tube on the
interlocking part.
6. Press the I.V Set firmly on the air detector
to fix the tube.
7. Grab the detector for Downstream
Occlusion by hand and firmly press it to fix
the tube.
Injection sets from right to left.
8. Insert the slide clip in the slide groove.
Then close the door.
Use of a clip is optional.