Alarms
and Actions
52
7.10
Low battery alarm
Item
Description
Cause
When the product is run on battery and the remaining operation time is 30
minutes before stop, a low battery alarm occurs.
Indicators
Alarm sounds
Alarm screen appears in yellow on the
LCD display.
LCD state display: “Low Battery”
Alarm clearing
method
When external power is supplied, the low battery alarm is cleared.
Press any buttons to clear the low battery alarm.
Action
Charge the battery by connecting to the AC power.
7.11
AC/DC removal power alarm
Item
Description
Cause
When the AC/DC power is disconnected, the AC/DC removal power alarm
occurs.
Indicators
Alarm sounds.
Alarm screen appears in yellow on the
LCD display.
LCD state display: “Removal Power”
Alarm clearing
method
When the AC/DC power is connected while in AC/DC removal power state,
the alarm is cleared or press any buttons to clear.
Action
Connect the AC/DC power.
7.12
K.V.O alarm
Item
Description
Cause
When the flow volume reaches the set amount, the injection is maintained
with K.V.O. F.Rate set to maintain the vein opening state.