Operation
32
5.3
Infusion start (operation started) and stop
Press the
START
button on the front of the product to start operation.
Infusion animation operates on the LCD screen, and infusion LED blinks sequentially.
When you press the START/STOP button while in normal operation (injecting or BOLUS
injection in progress), injection is stopped and the infusion animation and the green LED
turns off.
As operation progresses, injection volume increases on the I.Vol display at bottom of LCD.
Use the JOG button to check information about Drug, R.Time, T.Vol, I.Vol, F.Rate, and I.V
Set.
If you press the BOLUS button for 2~3 seconds, BOLUS is injected with the preset flow rate
and volume.
See
5.4 Operation Status
If you press the JOG button while in injection, you can change the flow rate.
See
5.4 Operation Status (Titration function).
When the current injected volume is within 5 minutes until total amount, the Near
Complete alarm is displayed.
See
6.4 Device Setting (Near complete alarm).
Press the
STOP
button to stop injection.
Infusion animation stop operating on the LCD screen, and infusion LED stop blinking.
Caution
Before injection, check if the I.V Set is mounted and the flow rate and the
volume to be injected are properly set.
After pressing the
START
button, check whether injection is made smoothly
for a specific time.
5.4
Operation status
5.4.1
Initializing Infusion Information and I.Vol
Operation Status
Description
Display
Initializing Infusion I
nformation
Clear Therapy
screen appears
by pressing the
CLEAR/BACK
button.
Initializing I.Vol
Use the
JOG
to select
I.Vol
and
press the
CLEAR/BACK
button.