Injection
Modes
26
4. Use the JOG to adjust the value of
Concentration, Weight, Dose, and T.Vol and
press the JOG to set ok.
The setting range of Weight is 0.1-300kg.
It is set as 0.1-99.9 kg (step of 0.1 kg), 1-300 kg
(step of 1 kg).
5. Select the Confirm at upper right corner and
press the JOG to apply the setting.
Then it returns to the Standby screen.
6. Press the START button to start injection.
4.5
Secondary injection mode
The secondary injection mode is required to run two different modes with different
settings including the primary injection mode. The modes can be run individually or
automatically.
1. Press the MENU button in the injection
standby screen.
Then the MENU screen appears.
Select
Infusion Setting.