Operation
33
5.4.2
Bolus (Purge) Injection Mode
Operation Status
Description
Display
Bolus
Emergency injection during the infusion
Purge
Emergency injection during the standby status
Auto Bolus
A preset pop-up window appears
by pressing the
BOLUS
button.
Auto Bolus
Select
Start
button to start Bolus
injection or
Modify
to adjust
F.Rate and T.Vol.
Manual Bolus
Bolus injection starts by pressing
the
BOLUS
button for 2 seconds
during infusion.
T.Vol
appears on
the pop-up screen.
Manual Bolus
Bolus injection stops and infusion
mode starts by unpressing the
BOLUS
button.
Caution
If you press the
BOLUS
button in standby mode, it detects the occlusion as a
purged status.
Be careful because overdose can happen momentarily.
Air sensor doesn’t function during
Purge Injection Mode
.
5.4.3
Titration-normal injection mode
1. Press the JOG for 2 seconds during injection.
Then the flow rate entry screen appears.
2. Use the JOG dial or MENU(left)/MUTE(right)
button to set the desired flow rate.
Press the
CLEAR
button to initialize the F.Rate (m/h)
back to ‘0’.