Filac

 

 3000 EZ/EZA Electronic Thermometer

6

Table of Contents

Direct Mode

1.   The Filac 3000 EZ/EZA Electronic Thermometer normally operates in Predictive mode for accurate temperature measuring.  In 

certain cases, such as with a hypothermic patient, the thermometer will automatically shift to Direct (slow) mode and will then 
act as a temperature monitor.  

2.   The Filac 3000 EZ/EZA Electronic Thermometer can be set to operate exclusively in Direct mode (disable Predictive mode).  See 

the Direct Mode Lock Section for instructions on configuring the thermometer to operate exclusively in Direct mode.

3.   In Direct mode the device may require up to 60 seconds  to reach equilibrium and display patient temperature.
4.  A turtle icon will be continuously displayed whenever the thermometer is functioning in Direct mode.
5.  To change between Celsius and Fahrenheit scales, press and release the °C/°F button.  Press and release again as needed.
6.   An Up or Down arrow will appear on the display whenever the current temperature reading is out of range, either High or 

Low, respectively.

7.   The Direct mode auto feature is always functional regardless of Red or Blue isolation chamber/ probe.
8.  A “long beep” is sounded when measurement is complete and the final temperature is displayed.
9.   After returning the probe to the probe well, the temperature is stored for recall until the probe is once again withdrawn.  If the 

probe is returned to the probe well before the “long beep” is heard, no temperature will be stored for the recall function.

Direct Mode Lock

1.   To configure the thermometer so it always operates in Direct mode, press and hold the Site and °C/°F buttons at the same time 

for 4 seconds.  A long “beep” will be sounded. On the LCD, the turtle will flash and “ON” will be displayed for approximately 
2 seconds before the unit turns off. (

Note:

 The thermometer must be off before configuring Direct mode lock.)

2.   In this mode of operation, the turtle icon will be continuously displayed during use to indicate Direct mode operation.
3.   To exit Direct mode lock and return to normal operation, again press and hold the Site and °C/°F buttons at the same time for 

4 seconds.  A long “beep” will be sounded. On the LCD, the turtle will flash and “OFF” will be displayed for approximately 
2 seconds then the unit will turn off.

Recall Last Temperature

1.   After each temperature measurement, a “long beep” is heard. The “beep” indicates the temperature measurement has been 

completed and stored and is available for recall. This temperature can be recalled after the probe is returned to the probe well.

2.   To recall the most recent temperature measurement, press and release °C/°F button on the front panel. The last measurement 

taken will appear for several seconds.

3.   While the recalled measurement is displayed, the user may press and release the °C/°F button again to change between the 

Celsius and Fahrenheit scales.

4.   Withdrawing the probe from the probe well erases last temperature memory.
5.   If the most recent temperature measurement was incomplete or out of range, dashes will appear on the display during the 

recall operation.

VI.  Cleaning and Sterilization

1.   The entire device may be easily wiped clean.  Water temperature should not exceed 130° F (55° C).  

Do not submerge or soak under water.

2.   A mild detergent may be added to water.  Use of cleaners such as Spray Nine

TM

*, pHisoHex

TM

*, Hibiclens

TM

*,

 

Vesta-Syde

TM

*, or Cidex

TM

* may result in damage to the thermometer case.

3.   Use of 10:1 water and hypoclorite mixture or a damp isopropyl alcohol wipe occasionally, is acceptable.   

Prolonged and repeated use of these chemicals may result in damage to the thermometer case and display area.

4.   Use of a cloth or sponge is recommended for cleaning.  Abrasive pads may result in damage to the thermometer case and 

display area.

5.   This thermometer is provided non-sterile.  

Do not

 use ethylene oxide gas, heat, autoclave, or any other harsh methods to 

sterilize this unit.

6.   Isolation chambers may be replaced inexpensively instead of cleaning.  
7.   After cleaning the unit, shake the probe handle to drain out any excess solution.  

Thoroughly dry the electrical contacts on both probe and thermometer. 

Summary of Contents for Filac 3000 EZ

Page 1: ...Table of Contents Filac TM 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 2: ...Filac 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 3: ...ectal Temperature Taking 5 Direct Mode 6 Direct Mode Lock 6 Recall Last Temperature 6 VI Cleaning and Sterilization 6 VII Battery Replacement Installation 7 VIII Troubleshooting 7 IX Specifications 8...

Page 4: ...urement and Regulation Plenum Press NY 1982 p 81 87 II General Information 1 The Filac 3000 EZ EZA Electronic Thermometer is a fast highly accurate and easy to use clinical instrument for measuring pa...

Page 5: ...not open unit No user serviceable parts inside Opening of device may affect calibration and voids warranty 11 Disposal of used probe covers must be performed in accordance with current medical practic...

Page 6: ...limitations 10 95 Temperature limitations 13 F 25 C 131 F 55 C Follow instructions for use This manual must be read prior to use of the thermometer Symbol appears blue on device Non ionizing electroma...

Page 7: ...mbersand Probes 1 For aiding in infection control use only the Blue probe and Blue isolation chamber for Oral and Axillary temperature taking The Red probe and Red isolation chamber must only be used...

Page 8: ...ss again as needed 13 Eject the used cover into a bio waste container by pushing top button 14 After returning the probe to the probe well the temperature is stored for recall until the probe is once...

Page 9: ...ring use to indicate Direct mode operation 3 To exit Direct mode lock and return to normal operation again press and hold the Site and C F buttons at the same time for 4 seconds A long beep will be so...

Page 10: ...top half of the door upwards pressing firmly on the top of the door until the small tab on the main unit clicks into the catch on the battery door 7 Reinstall isolation chamber with probe attached VII...

Page 11: ...ectal Mode 10 14 seconds Direct Mode All Sites 60 120 seconds Typical times for switch to Direct Mode If no measurement site detected 60 seconds after pulling probe from probe well If temperature does...

Page 12: ...ontinuous Use Intended for hand held use Degree of Safety of Application in the Presence Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Transport...

Page 13: ...mperature A Filac 3000 EZ EZA Electronic Thermometer should have a quick predict mode accuracy within 0 3 C of the calibrated water bath temperature The probe cable connector is designed to exceed 1 0...

Page 14: ...A Reorder No 505003 Complete Rectal Thermometer Includes electronic thermometer with Red Rectal isolation chamber and Red Rectal probe with 4 foot 1 2 m cord and global manual on CD Filac 3000 EZ Reor...

Page 15: ...ty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling Section XII Electromagnetic Conformity Declaration The Filac 3000 Ele...

Page 16: ...ions it is recommended that the P STIM be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 per EN 60601 1 2 2001 3 A m 3 A m Power frequen...

Page 17: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Page 18: ...U S and internationally registered trademarks of Covidien AG TM Trademark of its respective owner Other brands are trademarks of a Covidien company 2012 Covidien Made in China Covidien llc 15 Hampshir...

Reviews: