Device Materials:
Thermometer Case:
Flame retardant Polycarbonate/Polyester Blend
Isolation Chamber:
Flame retardant Polycarbonate/Polyester Blend
Probe Handle:
Flame retardant Polycarbonate/Polyester Blend
Probe Shaft:
Flame retardant Polyester
Probe Cable:
Polyurethane jacket with TPE over mold
Tip:
Aluminum
All materials are latex-free.
Type of Protection Against Electrical Shock:
Internally Powered Equipment. 4 AA Batteries.
Degree of Protection Against Electrical Shock:
Type BF
Degree of Protection Against Ingress of Fluids:
Not Protected-IPX0
Mode of Operation:
Continuous Use. Intended for hand held use.
Degree of Safety of Application in the Presence:
Not suitable for use in the presence of flammable anesthetic mixture with air,
oxygen or nitrous oxide.
Transport and Storage:
Not to exceed 30 days
Transport and store this device between temperatures of
-25°C to 55°C (-13°F to131°F).
Relative humidity range should be between 10%-95% RH, non-condensing.
Return unit to the factory if the device is dropped and performance appears degraded, or in the event that the environmental
conditions for transport, storage and operation are exceeded.
The Filac 3000 EZ/EZA Electronic Thermometer is in compliance with the following International Regulatory and
Safety standards:
UL60601-1
CAN/CSA C22.2 No. 60601-1-08
EN60601-1:2005
EN60601-1-2:2001
EN 12470-3:2000
MDD 93/42/EEC
EN980:2008
ASTM E 1112: 2006