 3000 EZ/EZA Electronic Thermometer

13

Table of Contents

Guidance and manufacturer’s declaration – electromagnetic immunity

The Filac 3000 Electronic Thermometer is intended for use in the electromagnetic environment specified below. The user of the Filac 
3000 Electronic Thermometer should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level Electromagnetic environment guidance

Electrostatic discharge 
(ESD) (EN 61000-4-2 per 
EN 60601-1-2: 2001)

± 6 kV contact 
± 8 kV air

± 6 kV contact 
± 8 kV air

Floors should be wood, concrete or ceramic tile. 
If floors are covered with synthetic material, the 
relative humidity should be at least 30 %.

Electrical fast transient / 
burst IEC 61000-4-4

± 2 kV for power supply lines
± 1 kV for input/output lines

not applicable
not applicable

Mains power quality should be that of a typical 
commercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV differential mode
± 2 kV common mode

not applicable
not applicable

Mains power quality should be that of a typical 
commercial or hospital environment.

Voltage dips, short 
interruptions and voltage 
variations on power 
supply input lines IEC 
61000-4-11

< 5 % UT (>95 % dip in UT ) 
for 0,5 cycle
40 % UT (60 % dip in UT ) 
for 5 cycles
70 % UT (30 % dip in UT ) 
for 25 cycles
< 5 % UT (>95 % dip in UT ) 
for 5 sec

not applicable

Mains power quality should be that of a typical 
commercial or hospital environment. If the 
user of the Filac 3000 Electronic Thermometer 
requires continued operation during power mains 
interruptions, it is recommended that the P-STIM be 
powered from an uninterruptible power supply or 
a battery.

Power frequency (50/60 
Hz) magnetic field (EN 
61000-4-8 per EN 60601-
1-2: 2001)

3 A/m

3 A/m

Power frequency magnetic fields should be at 
levels characteristic of a typical location in a typical 
commercial or hospital environment.

Note

 UT is the a. c. mains voltage prior to application of the test level.

Recommended separation distances between portable and mobile RF communications equipment 

and the Filac 3000 Electronic Thermometer

The Filac 3000 Electronic Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances 
are controlled. The customer or the user of the Filac 3000 Electronic Thermometer can help prevent electromagnetic interference by 
maintaining the minimum distance between portable and mobile RF communications
Rated maximum output power 
of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz 

d= 1,2√P

80 MHz to 800 MHz 

d=1,2√P

800 MHz to 2,5 GHz 

d = 1,2√P

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

2

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be 
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter manufacturer.

Note 1

 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflections from 

structures, objects and people.

Summary of Contents for Filac 3000 EZ

Page 1: ...Table of Contents Filac TM 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 2: ...Filac 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 3: ...ectal Temperature Taking 5 Direct Mode 6 Direct Mode Lock 6 Recall Last Temperature 6 VI Cleaning and Sterilization 6 VII Battery Replacement Installation 7 VIII Troubleshooting 7 IX Specifications 8...

Page 4: ...urement and Regulation Plenum Press NY 1982 p 81 87 II General Information 1 The Filac 3000 EZ EZA Electronic Thermometer is a fast highly accurate and easy to use clinical instrument for measuring pa...

Page 5: ...not open unit No user serviceable parts inside Opening of device may affect calibration and voids warranty 11 Disposal of used probe covers must be performed in accordance with current medical practic...

Page 6: ...limitations 10 95 Temperature limitations 13 F 25 C 131 F 55 C Follow instructions for use This manual must be read prior to use of the thermometer Symbol appears blue on device Non ionizing electroma...

Page 7: ...mbersand Probes 1 For aiding in infection control use only the Blue probe and Blue isolation chamber for Oral and Axillary temperature taking The Red probe and Red isolation chamber must only be used...

Page 8: ...ss again as needed 13 Eject the used cover into a bio waste container by pushing top button 14 After returning the probe to the probe well the temperature is stored for recall until the probe is once...

Page 9: ...ring use to indicate Direct mode operation 3 To exit Direct mode lock and return to normal operation again press and hold the Site and C F buttons at the same time for 4 seconds A long beep will be so...

Page 10: ...top half of the door upwards pressing firmly on the top of the door until the small tab on the main unit clicks into the catch on the battery door 7 Reinstall isolation chamber with probe attached VII...

Page 11: ...ectal Mode 10 14 seconds Direct Mode All Sites 60 120 seconds Typical times for switch to Direct Mode If no measurement site detected 60 seconds after pulling probe from probe well If temperature does...

Page 12: ...ontinuous Use Intended for hand held use Degree of Safety of Application in the Presence Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Transport...

Page 13: ...mperature A Filac 3000 EZ EZA Electronic Thermometer should have a quick predict mode accuracy within 0 3 C of the calibrated water bath temperature The probe cable connector is designed to exceed 1 0...

Page 14: ...A Reorder No 505003 Complete Rectal Thermometer Includes electronic thermometer with Red Rectal isolation chamber and Red Rectal probe with 4 foot 1 2 m cord and global manual on CD Filac 3000 EZ Reor...

Page 15: ...ty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling Section XII Electromagnetic Conformity Declaration The Filac 3000 Ele...

Page 16: ...ions it is recommended that the P STIM be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 per EN 60601 1 2 2001 3 A m 3 A m Power frequen...

Page 17: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Page 18: ...U S and internationally registered trademarks of Covidien AG TM Trademark of its respective owner Other brands are trademarks of a Covidien company 2012 Covidien Made in China Covidien llc 15 Hampshir...

Reviews: