background image

Filac

 

 3000 EZ/EZA Electronic Thermometer

5

Table of Contents

Oral & Axillary Temperature Taking
  1.  Make certain that the Blue isolation chamber/probe unit is attached.
  2.  Withdraw probe and apply a probe cover.  
 

The thermometer turns on automatically.

  3.   An icon identifying  Oral  or Axillary mode is displayed.  The Rectal icon can not be 

displayed when a blue isolation chamber/probe is attached.

  4.  Press the Site button on the front panel to select either the  Oral  or Axillary mode.
  5.   For Oral temperatures, insert the probe tip deep into the sublingual pocket next 

to the frenulum linguae, (vertical fold of tissue in middle of tongue), on one side 

or the other, toward the back of the mouth.

Note:

 Accurate body temperature readings can only be obtained in one of these two 

“heat pocket” locations as shown.  Temperatures taken in other mouth locations will 

result in inaccurate body temperature readings.
  6.  Patient’s mouth must be 

closed

.*

  7.   Securely hold the probe in place until the temperature 

is displayed.

  8.   For Axillary temperatures, have the patient raise the arm, then 

place the probe tip in the axilla.  Press gently to assure good 

contact. For the most accurate temperature the probe tip 

should be placed directly against the patients skin.

  9.   Have the patient then lower the arm and remain as still as 

possible.  Hold the probe parallel to the arm as shown.*

10.   If two short beeps are heard, it means the unit switched to 

Direct (slow) mode for this temperature only.

11.   A “long beep” is sounded when measurement is complete and 

the final temperature is displayed.

12.   To change between Celsius and Fahrenheit scales, press the 

°C/°F button.  Press again as needed.

13.   Eject the used cover into a bio-waste container by pushing 

top button.

14.   After returning the probe to the probe well, the temperature is 

stored for recall until the probe is once again withdrawn.  If 

the probe is returned to the probe well before the “long beep” 

is heard, no temperature will be stored for the recall function.

Rectal Temperature Taking
  1.   Make certain that the Red isolation chamber/probe unit is attached.
  2.   Withdraw the probe and apply a probe cover.  Thermometer turns on automatically.
  3.   An icon identifying Rectal mode is always displayed provided the Red isolation chamber/probe assembly is attached.  

Pressing the Site button on the front panel to change modes has no effect.

  4.   If desired, apply medical grade lubricant as per institutional protocol.
  5.   Insert the probe into the patient’s rectum.  To ensure proper tissue contact, angle the probe slightly after insertion.*
  6.   Depth of insertion is recommended at 1/2” to 3/4” (12 mm - 19 mm) for adults and 1/4” to 1/2” (6 mm - 13 mm) 

for children.

  7.   If two short beeps are heard, it means the unit switched to Direct (slow) mode for this temperature only.
  8.   A “long beep” is sounded when measurement is complete and the final temperature is displayed.
  9.   To change between Celsius and Fahrenheit scales, press the °C/°F button.  Press again as needed.
10.  Eject the used cover into a bio-waste container by pushing top button.
11.   After returning the probe to the probe well, the temperature is stored for recall until the probe is once again 

withdrawn.  If the probe is returned to the probe well before the “long beep” is heard, no temperature will be stored 

for the recall function.

*Probe movement during a measurement can affect the thermometer’s ability to accurately measure the site temperature and 
may lengthen the time required to obtain a reading.

Heat Pockets

Summary of Contents for Filac 3000 EZ

Page 1: ...Table of Contents Filac TM 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 2: ...Filac 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 3: ...ectal Temperature Taking 5 Direct Mode 6 Direct Mode Lock 6 Recall Last Temperature 6 VI Cleaning and Sterilization 6 VII Battery Replacement Installation 7 VIII Troubleshooting 7 IX Specifications 8...

Page 4: ...urement and Regulation Plenum Press NY 1982 p 81 87 II General Information 1 The Filac 3000 EZ EZA Electronic Thermometer is a fast highly accurate and easy to use clinical instrument for measuring pa...

Page 5: ...not open unit No user serviceable parts inside Opening of device may affect calibration and voids warranty 11 Disposal of used probe covers must be performed in accordance with current medical practic...

Page 6: ...limitations 10 95 Temperature limitations 13 F 25 C 131 F 55 C Follow instructions for use This manual must be read prior to use of the thermometer Symbol appears blue on device Non ionizing electroma...

Page 7: ...mbersand Probes 1 For aiding in infection control use only the Blue probe and Blue isolation chamber for Oral and Axillary temperature taking The Red probe and Red isolation chamber must only be used...

Page 8: ...ss again as needed 13 Eject the used cover into a bio waste container by pushing top button 14 After returning the probe to the probe well the temperature is stored for recall until the probe is once...

Page 9: ...ring use to indicate Direct mode operation 3 To exit Direct mode lock and return to normal operation again press and hold the Site and C F buttons at the same time for 4 seconds A long beep will be so...

Page 10: ...top half of the door upwards pressing firmly on the top of the door until the small tab on the main unit clicks into the catch on the battery door 7 Reinstall isolation chamber with probe attached VII...

Page 11: ...ectal Mode 10 14 seconds Direct Mode All Sites 60 120 seconds Typical times for switch to Direct Mode If no measurement site detected 60 seconds after pulling probe from probe well If temperature does...

Page 12: ...ontinuous Use Intended for hand held use Degree of Safety of Application in the Presence Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Transport...

Page 13: ...mperature A Filac 3000 EZ EZA Electronic Thermometer should have a quick predict mode accuracy within 0 3 C of the calibrated water bath temperature The probe cable connector is designed to exceed 1 0...

Page 14: ...A Reorder No 505003 Complete Rectal Thermometer Includes electronic thermometer with Red Rectal isolation chamber and Red Rectal probe with 4 foot 1 2 m cord and global manual on CD Filac 3000 EZ Reor...

Page 15: ...ty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling Section XII Electromagnetic Conformity Declaration The Filac 3000 Ele...

Page 16: ...ions it is recommended that the P STIM be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 per EN 60601 1 2 2001 3 A m 3 A m Power frequen...

Page 17: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Page 18: ...U S and internationally registered trademarks of Covidien AG TM Trademark of its respective owner Other brands are trademarks of a Covidien company 2012 Covidien Made in China Covidien llc 15 Hampshir...

Reviews: