Filac

 

 3000 EZ/EZA Electronic Thermometer

2

Table of Contents

III.  Safety and Warnings

Caution:  Federal law (USA) restricts this device to sale by or on the order of a physician.

  1.   Read this booklet thoroughly before using the Filac 3000 EZ/EZA Electronic Thermometer.
  2.   Do not use this device near flammable anesthetics.  Not suitable for use in the presence of flammable anesthetic mixture with 

air, oxygen or nitrous oxide, or an oxygen rich environment.

  3.   Do not use this thermometer without first installing a new Filac 3000 EZ/EZA Electronic Thermometer probe cover.
  4.   Use only Filac 3000 EZ/EZA Electronic Thermometer probe covers

 

with

 

this device.

 

Use of any other probe cover will result in erroneous temperature readings.

  5.   The device and probe covers are Non-sterile.  Do not use on abraded tissue.
  6.   To limit cross contamination, use Blue devices for Oral and Axillary temperature taking only.
  7.  Use 

red

 devices only for 

rectal

 temperatures.

  8.  Thoroughly dry all electrical contacts on both probe and thermometer after washing, or device may fail to function properly.
  9.  For re-calibration, service or integrity checks, refer to a qualified Biomedical Technician or return to manufacturer.
10.   

Warning:

 No modification of this equipment is allowed. Do not open unit.  No user-serviceable parts inside.  Opening of device 

may affect calibration and voids warranty.

11.   Disposal of used probe covers must be performed in accordance with current medical practices or local regulations regarding 

disposal of infectious, biological medical waste.

12.   Do not use lithium batteries with this device.
13.   Do not mix alkaline and rechargeable batteries.
14.   Removal of the batteries is recommended if the unit is not going to be used for an extended period of time.
15.   Dispose of batteries in a manner consistent with local environmental and institutional policy for Alkaline battery disposal.
16.   Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired 

equipment disposal.

17.   Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices. 
18.   Device to be used by trained personnel.

Note:

Even though this device has been designed to minimize the effects of electromagnetic interference, it does generate radio frequency 
energy.  If not used in accordance with the instructions, the device could cause interference in other equipment operating within its 
vicinity.  If the device is causing interference, the following actions may be taken in an attempt to correct the interference:

• 

Re-orient or re-locate the receiving device.

• 

Increase the separation between the devices. 

• 

Consult a customer service representative.

Summary of Contents for Filac 3000 EZ

Page 1: ...Table of Contents Filac TM 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 2: ...Filac 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 3: ...ectal Temperature Taking 5 Direct Mode 6 Direct Mode Lock 6 Recall Last Temperature 6 VI Cleaning and Sterilization 6 VII Battery Replacement Installation 7 VIII Troubleshooting 7 IX Specifications 8...

Page 4: ...urement and Regulation Plenum Press NY 1982 p 81 87 II General Information 1 The Filac 3000 EZ EZA Electronic Thermometer is a fast highly accurate and easy to use clinical instrument for measuring pa...

Page 5: ...not open unit No user serviceable parts inside Opening of device may affect calibration and voids warranty 11 Disposal of used probe covers must be performed in accordance with current medical practic...

Page 6: ...limitations 10 95 Temperature limitations 13 F 25 C 131 F 55 C Follow instructions for use This manual must be read prior to use of the thermometer Symbol appears blue on device Non ionizing electroma...

Page 7: ...mbersand Probes 1 For aiding in infection control use only the Blue probe and Blue isolation chamber for Oral and Axillary temperature taking The Red probe and Red isolation chamber must only be used...

Page 8: ...ss again as needed 13 Eject the used cover into a bio waste container by pushing top button 14 After returning the probe to the probe well the temperature is stored for recall until the probe is once...

Page 9: ...ring use to indicate Direct mode operation 3 To exit Direct mode lock and return to normal operation again press and hold the Site and C F buttons at the same time for 4 seconds A long beep will be so...

Page 10: ...top half of the door upwards pressing firmly on the top of the door until the small tab on the main unit clicks into the catch on the battery door 7 Reinstall isolation chamber with probe attached VII...

Page 11: ...ectal Mode 10 14 seconds Direct Mode All Sites 60 120 seconds Typical times for switch to Direct Mode If no measurement site detected 60 seconds after pulling probe from probe well If temperature does...

Page 12: ...ontinuous Use Intended for hand held use Degree of Safety of Application in the Presence Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Transport...

Page 13: ...mperature A Filac 3000 EZ EZA Electronic Thermometer should have a quick predict mode accuracy within 0 3 C of the calibrated water bath temperature The probe cable connector is designed to exceed 1 0...

Page 14: ...A Reorder No 505003 Complete Rectal Thermometer Includes electronic thermometer with Red Rectal isolation chamber and Red Rectal probe with 4 foot 1 2 m cord and global manual on CD Filac 3000 EZ Reor...

Page 15: ...ty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling Section XII Electromagnetic Conformity Declaration The Filac 3000 Ele...

Page 16: ...ions it is recommended that the P STIM be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 per EN 60601 1 2 2001 3 A m 3 A m Power frequen...

Page 17: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Page 18: ...U S and internationally registered trademarks of Covidien AG TM Trademark of its respective owner Other brands are trademarks of a Covidien company 2012 Covidien Made in China Covidien llc 15 Hampshir...

Reviews: