VII.  Battery Replacement / Installation

 

1.   To access the batteries, remove the isolation chamber by depressing the tabs on 

the sides of the unit and slide off upwardly.

 

2.   Pull back on the finger tab on the top of the battery door.  The battery door should 

hinge open to expose the batteries. 

 

3.   The batteries may be removed by pressing firmly on the positive end (opposite the 

compressed spring) and lifting up.

 

4.   Be certain to dispose of used batteries in a manner consistent with local 

environmental and institutional policy for lithium or alkaline battery disposal.

 

5.   To install new batteries, place the negative end on the spring.  Press down firmly to 

compress spring and tilt battery into place.

 

6.   To replace the battery door, insert the tab on the bottom of the door into the slot 

on the bottom of the unit to create a hinge.  Swing the top half of the door 
upwards pressing firmly on the top of the door until the small tab on the main unit 
“clicks” into the catch on the battery door.

 

7.   Reinstall isolation chamber with probe attached.

VIII.  Troubleshooting

1.   If device fails to function properly after washing & reconnecting probe, rinse all 

contacts in de-ionized water and blow dry.

2.  Icons indicate all other possible errors and remedies.
3.  Device Error Codes:
 

E01 - System error during synchronization 

 

E02 - System error during patient thermistor calibration

 

E03 - System error during heater thermistor calibration

 

E04 - System timing error

 

E05 - Heater error

 

E06 - LCD communication error

 

E07 - Shut off error

 

P01 - Probe configuration (or no probe connected) error

 

P02 - Direct Mode patient thermistor unstable or out of range 

 

P03 - Direct Mode heater thermistor unstable or out of range

 

P04 - Predict Mode patient thermistor unstable or out of range 

 

P05 - Predict Mode heater thermistor unstable or out of range

 

P06 - Unable to pre-heat probe tip

4.  No user-serviceable parts inside.
5.  Refer to a factory or qualified Biomedical Technician for service.

Summary of Contents for Filac 3000 EZ

Page 1: ...Table of Contents Filac TM 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 2: ...Filac 3000 EZ EZA Electronic Thermometer Operating Manual...

Page 3: ...ectal Temperature Taking 5 Direct Mode 6 Direct Mode Lock 6 Recall Last Temperature 6 VI Cleaning and Sterilization 6 VII Battery Replacement Installation 7 VIII Troubleshooting 7 IX Specifications 8...

Page 4: ...urement and Regulation Plenum Press NY 1982 p 81 87 II General Information 1 The Filac 3000 EZ EZA Electronic Thermometer is a fast highly accurate and easy to use clinical instrument for measuring pa...

Page 5: ...not open unit No user serviceable parts inside Opening of device may affect calibration and voids warranty 11 Disposal of used probe covers must be performed in accordance with current medical practic...

Page 6: ...limitations 10 95 Temperature limitations 13 F 25 C 131 F 55 C Follow instructions for use This manual must be read prior to use of the thermometer Symbol appears blue on device Non ionizing electroma...

Page 7: ...mbersand Probes 1 For aiding in infection control use only the Blue probe and Blue isolation chamber for Oral and Axillary temperature taking The Red probe and Red isolation chamber must only be used...

Page 8: ...ss again as needed 13 Eject the used cover into a bio waste container by pushing top button 14 After returning the probe to the probe well the temperature is stored for recall until the probe is once...

Page 9: ...ring use to indicate Direct mode operation 3 To exit Direct mode lock and return to normal operation again press and hold the Site and C F buttons at the same time for 4 seconds A long beep will be so...

Page 10: ...top half of the door upwards pressing firmly on the top of the door until the small tab on the main unit clicks into the catch on the battery door 7 Reinstall isolation chamber with probe attached VII...

Page 11: ...ectal Mode 10 14 seconds Direct Mode All Sites 60 120 seconds Typical times for switch to Direct Mode If no measurement site detected 60 seconds after pulling probe from probe well If temperature does...

Page 12: ...ontinuous Use Intended for hand held use Degree of Safety of Application in the Presence Not suitable for use in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Transport...

Page 13: ...mperature A Filac 3000 EZ EZA Electronic Thermometer should have a quick predict mode accuracy within 0 3 C of the calibrated water bath temperature The probe cable connector is designed to exceed 1 0...

Page 14: ...A Reorder No 505003 Complete Rectal Thermometer Includes electronic thermometer with Red Rectal isolation chamber and Red Rectal probe with 4 foot 1 2 m cord and global manual on CD Filac 3000 EZ Reor...

Page 15: ...ty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling Section XII Electromagnetic Conformity Declaration The Filac 3000 Ele...

Page 16: ...ions it is recommended that the P STIM be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 per EN 60601 1 2 2001 3 A m 3 A m Power frequen...

Page 17: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Page 18: ...U S and internationally registered trademarks of Covidien AG TM Trademark of its respective owner Other brands are trademarks of a Covidien company 2012 Covidien Made in China Covidien llc 15 Hampshir...

Reviews: