background image

 

HH806U/HH806AU 

MULTILOGGER THERMOMETER

 

 
 
 
 
 
 
 
 
 

 
 
 
 

 

 

 

OMEGAnet

®

 On-Line Service

omega.com

 

Internet e-mail 

[email protected]

 

Servicing North America: 

U.S.A:

 

One Omega Drive, Box 4047 

 

Stamford, CT 06907-0047 

ISO 9001

  Tel: (203) 359-1660 

FAX: (203) 359-7700

 

Certified

 e-mail: 

[email protected] 

Canada:

 976 

Bergar 

 

Laval (Quebec) H7L 5A1, Canada 

 

Tel: (514) 856-6928 

FAX: (514) 856-6886 

 e-mail: 

[email protected] 

 

For immediate technical or application assistance: 

U.S.A 

Sales Service: 1-800-826-6342/1-800-TC-OMEGA

® 

and 

Customer Service: 1-800-622-2378/1-800-622-BEST

® 

Canada: 

Engineering Service: 1-800-872-9436/1-800-USA-WHEN

®

 

 

TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA

 

Mexico:

 En 

Espan

ol: (001) 203-359-7803 

 e-mail: 

[email protected] 

 

FAX: (001) 203-359-7807  [email protected] 

 

Servicing Europe: 

Benelux: 

Postbus 8034, 1180 LA Amstelveen, The Netherlands 

 

Tel: +31 (0)20 3472121 

FAX: +31 (0)20 6434643 

 

Toll Free in Benelux: 0800 0993344 

 e-mail: 

[email protected]

 

Czech

 

Frystatska 184, 733 01 Karviná, Czech Republic 

Republic: 

Tel: +420 (0)59 6311899 

FAX: +420 (0)59 6311114 

 

Toll Free: 0800-1-66342 

e-mail: [email protected] 

France:

 

11, rue Jacques Cartier, 78280 Guyancourt, France 

 

Tel: +33 (0)1 61 37 2900 

FAX: +33 (0)1 30 57 5427 

 

Toll Free in France: 0800 466 342 

 e-mail: 

[email protected] 

Germany/

  Daimlerstrasse 26, D-75392 Deckenpfronn, Germany 

Austria:

 

Tel: + 49 (0)7056 9398-0  FAX: +49 (0)7056 9398-29 

 

Toll Free in Germany: 0800 639 7678 

 e-mail: 

[email protected] 

United

 

One Omega Drive, River Bend Technology Centre 

Kingdom:

  Northbank, Irlam, Manchester 

ISO 9002

  M44 5BD United Kingdom 

Certified

 

Tel: +44 (0)161 777 6611  FAX: +44 (0)161 777 6622 

 

Toll Free in United Kingdom: 0800-488-488 

 e-mail: 

[email protected] 

 
 
 
 

It is the policy of OMEGA Engineering, Inc. to comply with all 
worldwide safety and EMC/EMI regulations that apply. OMEGA 
is constantly pursuing certification of its products to the 
European New Approach Directives. OMEGA will add the CE 

mark to every appropriate device upon certification. 
The information contained in this document is believed to be 
correct, but OMEGA accepts no liability for any errors it 
contains, and reserves the right to alter specifications without 
notice. 

WARNING:

 These products are not designed for use in, and 

should not be used for, human applications. 

INTRODUCTION

 

The instrument is a portable digital thermometer that 

measures external thermocouples of type K, J, R, S, T, E, 
N. The thermocouples types comply with the N.I.S.T. – 
ITS 90 standard reference temperature/voltage tables. 
The thermometer features a dual thermocouple input, an 
adjustable T/C offset and an USB interface for uploading 
data to a PC using optional software and cable.

 

 

SAFETY INFORMATION 

WARNING 

To avoid electrical shock, do not use this instrument 

when working voltages at the measurement surface 

over 24V AC or DC.

 

 

WARNING 

To avoid damage or burns, do not make temperature 

measurement in microwave ovens.

 

 

CAUTION 

Repeated sharp flexing can break the thermocouple 

leads. To prolong lead life, avoid sharp bends in the 

leads, especially near the connector.

 

 

SPECIFICATIONS 

ELECTRICAL 

Temperature Scale:

 Celsius or Fahrenheit user-selectable 

Measurement Range: 

K-TYPE (0.1°)  -200°C to 1372°C (-328°F to 2501°F) 
J-TYPE (0.1°)  -210°C to 1200°C ( -346°F to 2192°F) 
T-TYPE (0.1°)  -200°C to 400°C (-328°F to 752°F) 
E-TYPE (0.1°)  -210°C to 1000°C (-346°F to 1832°F) 
R-TYPE (1°) 

0°C to 1767°C (32°F to 3212°F) 

S-TYPE (1°) 

0°C to 1767°C ( 32°F to 3212°F) 

N-TYPE (0.1°)  -50°C to 1300°C ( -58°F to 2372°F) 

*Based on the ITS-90 temperature standard. 

According to temperature standard ITS-90. 

Accuracy: 

K/J/T/E-TYPE 

±(0.05% rdg + 0.3°C) on -50°C to 1372°C 
±(0.05% rdg + 0.7°C) on -50°C to -210°C 
±(0.05% rdg + 0.6°F) on -58°F to 2501°F 
±(0.05% rdg + 1.4°F) on -58°F to -346°F 

N-TYPE 

±(0.05% rdg + 0.8°C) on -50°C to 0°C 
±(0.05% rdg + 0.4°C) on 0°C to 1300°C 
±(0.05% rdg + 1.6°F) on -58°F to 32°F 
±(0.05% rdg + 0.8°F) on 32°F to 2372°F 

R/S-TYPE 

±(0.05% rdg + 2°C) on 0°C to 1767°C 
±(0.05% rdg + 4°F) on 32°F to 3212°F 

Temperature Coefficient: 

0.1 times the applicable accuracy specification per °C 
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 
82°F to122°F). 
 

Input Protection: 

24V dc or 24V ac rms maximum input voltage on any 
combination of input pins. 

Reading Rate: 

2.5 time per second. 

 

ENVIRONMENTAL 

Ambient Operating Ranges: 

0°C to 50°C (32°F to 122°F) <80% R.H. 

Storage Temperature: 

-20°C to 60°C (-4°F to 140°F) <70% R.H. 
 

GENERAL 

Display:  

There are three display areas on the HH806 series LCD 
(liquid crystal display). The Main and Second displays 
are 4 ½ digits with a maximum reading of 19999. These 
are used for displaying the value of T1 or T2. The Third 
display is for the date, time, or the differential value of 
T1 to T2.

 

Overload: 

“----.-” or “OL” is display. 

Battery: 

1.5V x 4 PCS (SIZE AAA) UM-4 R03. 

Battery Life: 

120 hours typical with carbon zinc battery. 

Dimensions: 

160mm(H) x 83mm(W) x 38mm(D). 

Weight: 

Approx. 365g including batteries. 

Supplied Thermocouple: 

 

1 meter (40”) type K insulated beaded wire thermocouple. 
Maximum insulation temperature is 482°C (900°F). 
Thermocouple accuracy is ±1.1°C or 0.4% of reading 
(whichever is greater) from 0°C to 1250°C. 

Wire Communication Protocol:

   

2400 baud rate. (HH806U) 
19200 baud rate. (HH806A) 
 

External Connections: 

1. USB Port 
2. DC power JACK(12V) 

M4502/0707 

Shop online at: omega.com e-mail: [email protected] 

For latest product manuals: omegamanual.info 