![Copper Mountain Technologies PLANAR TR1300/1 Programming Manual Download Page 319](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189319.webp)
COM Server Commands
319
SCPI.TRIGger.SEQuence.SINGle
Object Type
Method
Target
Instrument
Description
Generates a trigger, independently of the trigger source setting (except
for the
External
). If the trigger source is set to
External
, an error occurs
and the command is ignored.
If the instrument is not in the waiting for a trigger state (sweep is in
progress or all the channels are set to
Hold
), an error occurs and the
command is ignored.
The method does not return control before the end of the sweep
(waiting for the completion of the sweep of all the channels).
Syntax
app.SCPI.TRIGger.SEQuence.SINGle
Related
Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent
Softkeys
None