COM Server Commands
184
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.ERESponse
Description
Selects the port numbers and sets the
1
–
path 2
–
port calibration
type,
when the written calibration coefficients are made effective by the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
method.
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Type
Variant: array of long (write only)
Range
Port number is 1–2. Array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs. If the same
port numbers are specified, an error occurs.
Target
Channel
Ch
,
Ch
Syntax
Ports
= Array(2, 1)
app
.SCPI.SENSe(
Ch
).CORRection.COEFficient.METHod.
ERESponse =
Ports
Equivalent
Softkeys
None