SCPI.SENSe(Ch).OFFSet.PORT(Pt).FREQuency.DIVisor
Object Type
Property (read/write)
Data Type
Double
Target
Port
Pt
of channel
Ch
,
Ch
Pt
port number 1–2 (see
Description
The basic frequency range divisor when the frequency offset feature is
ON and offset type is "PORT"..
Range
from 1 to 1000
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
n/a
Syntax
Dim Value As Double
Value =
app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.
DIVisor
app.SCPI.SENSe(Ch).OFFset.PORT(Pt).FREQuency.
DIVisor = 2
Related
Commands
SCPI.SENSe(Ch).OFFset.STATe
SCPI.SENSe(Ch).OFFset.TYPE
Equivalent
Softkeys
Stimulus > Freq.Offset > Port 1 | Port 2 > Divider