COM Server Commands
155
SCPI.MMEMory.LOAD.CKIT(
Ck
)
Object Type
Property (write only)
Data Type
String
Target
Calibration kit
Ck
,
Ck
calibration kit number 1–11 (see
Description
Recalls the definition file for the calibration kit. The file must be saved
by the
SCPI.MMEMory.STORe.CKIT(Ck
) property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.CKIT(Ck) = File
Notes
If the full path to the file is not specified, the \CalKit subdirectory of
the main directory will be searched for the file. The calibration kit
definition file has *.ckd extension by default.
Equivalent
Softkeys
None
Table 7.
Ck
: Calibration Kit Number
Data Type
Long
Description
Calibration kit number.
Range
from 1 to 16
Out of Range
An error occurs. Error code: 114.
Notes
If the calibration kit number is not specified, it is taken as equal to 1.