COM Server Commands
231
SCPI.SENSe(
Ch
).CORRection.EXTension.PORT(
Pt
).LOSS(
Ls
)
Object Type
Property (read/write)
Data Type
Double
Target
Port
Pt
of channel
Ch
,
Ch
Pt
port number 1–2 (see
Description
The loss value at the point number
Ls
to calculate the loss for the port
extension function,
Ls
Range
from -200 to 200
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB (decibel)
Syntax
Dim Value As Double
Value =
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).
LOSS(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).
LOSS(Ls) = 10
Equivalent
Softkeys
Calibration > Port Extensions > Loss > Loss 1 | Loss 2