COM Server Commands
226
SCPI.SENSe(
Ch
).CORRection.COLLect.METHod.TYPE
Object Type
Property (read only)
Data Type
String
Target
Channel
Ch
,
Ch
Description
The calibration type selected for calculating of the calibration
coefficients on completion of the calibration executed by the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE
method.
Range
"RESPO"
Response (Open)
"RESPS"
Response (Short)
"RESPT"
Response (Thru)
"SOLT1"
Full 1–port calibration
"1PATH"
One path 2–port calibration
"NONE"
Not defined
Syntax
Dim Param As String
Param =
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE
Equivalent
Softkeys
None