![Copper Mountain Technologies PLANAR TR1300/1 Programming Manual Download Page 191](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189191.webp)
COM Server Commands
191
SCPI.SENSe(
Ch
).CORRection.COLLect.ACQuire.SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel
Ch
,
Ch
Description
Measures the calibration data of the short standard for the specified
port.
Range
Port number is 1 or 2.
Out of Range
An error occurs.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
= 1
Notes
The property writing starts the measurement for the channel
independently of the trigger initiation and trigger source settings. The
function of the property writing waits for the completion of the
measurement.
Equivalent
Softkeys
Calibration > Calibrate > Response (Short) > Short
Calibration > Calibrate > Full 1–Port Cal > Short
Calibration > Calibrate > One Path 2–Port Cal > Short