![Copper Mountain Technologies PLANAR TR1300/1 Programming Manual Download Page 185](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189185.webp)
COM Server Commands
185
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.OPEN
Description
Selects the port and sets the response calibration (Open) type, when
the written calibration coefficients are made effective by the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
method.
Type
Long (write only)
Range
Port number is 1 or 2. (see Table 2)
Out of Range
An error occurs.
Target
Channel
Ch
,
Ch
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.
RESPonse.OPEN = Port
Equivalent
Softkeys
None
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.RESPonse.SHORt
Description
Selects the port and sets the response calibration (Short) type, when
the written calibration coefficients are made effective by the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE
method.
Type
Long (write only)
Range
Port number is 1 or 2. (see Table 2)
Out of Range
An error occurs.
Target
Channel
Ch
,
Ch
Syntax
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.
RESPonse.SHORt = Port
Equivalent
Softkeys